N5531S Measuring Receiver Performance Verification
Performance Verification Tests
Table 8-5
Example Test Record
Reference Attenuator Cal Data
S/N
Atten
Atten
Calibration
Uncert
Value
(± dB)
(dB)
0
0
9.989
0.006
20.042
0.006
30.025
0.006
40.194
0.006
50.175
0.011
370
Attenuator
Linearity Error
Measured
(dB)
(Attenuator Measured −
(dB)
Atten Cal value)
0
0
− 0.006
9.983
20.044
0.002
− 0.003
30.022
− 0.004
40.190
− 0.006
50.169
Specification
(± dB)
0
0.014
0.019
0.024
0.029
0.034