Agilent Technologies PSA Series Manual page 185

Spectrum analyzer measuring receiver system measuring receiver guide
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Preset:
State Saved:
Range:
Example:
4.4.3.2.3 Range Switching
Two kinds of range switching are supported. One is Manual ranging which is similar to what 8902A does. For
example, when you use 0dBm and –60 dBm signal to calibrate measuring receiver, while the signal is near 0
dBm and –60 dBm, recal status is displayed, then TRFL Calibrate key should be pressed. In Auto mode, the
TRFL Calibrate key is grayed out, and the range switching can be made automatically depending on SNR.
Key Path:
Remote Command:
Preset:
State Saved:
Range:
Dependencies/Couplings:
Example:
4.4.3.2.4 Range Hold
In TRFL measurement, there are 3 ranges which may be used to measure over a range of
>100 dB. This parameter forces the measurement to remain in the current range. One use for this parameter is
when you are measuring unstable signals, which can cause the analyzer to switch between 2 different ranges.
Selecting
holds the analyzer in the current range.
RangeHold
Mode:
Key Path:
Remote Command:
Preset:
State Saved:
Range:
Chapter 4
H10
Saved in instrument state.
H10|H75|H30k|H200k
:TRFL:IFBW H10
:TRFL:IFBW?
Meas Setup
[:SENSe]:TRFLevel:RASWitch AUTO|MAN
[:SENSe]:TRFLevel:RASWitch?
AUTO
Saved in instrument state.
Auto|Man
While Auto is selected, TRFL calibrated key must be grayed.
:TRFL:RASW MAN
:TRFL:RASW?
MRECEIVE
Meas Setup
[:SENSe]:TRFLevel:RHOLd OFF|ON|0|1
[:SENSe]:TRFLevel:RHOLd?
OFF
Saved in instrument state.
On | Off
Front-Panel Key and SCPI Command Reference
Measurement keys
185