Agilent Technologies PSA Series Manual page 157

Spectrum analyzer measuring receiver system measuring receiver guide
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

4.3.2 Frequency Counter
Allows the user to switch to the Frequency Counter measurement.
Mode:
Key Path:
Remote Command:
Notes:
Example:
4.3.2.1 SCPI Remote Commands
:CONFigure:FCOunter
:INITiate:FCOunter
:FETCh:FCOunter[n]?
:MEASure:FCOunter[n]?
:READ:FCOunter[n]?
Index: n
Results Returned
<Mnemonic>
not specified or
Returns the following 4 values:
n = 1
Frequency (Hz)
Frequency error (when set to Manual tune mode, 0 in Auto Tuning Mode)
Ratio Reference Frequency (Hz) (–999 in Normal mode, depends on Display Mode setting)
Frequency Ratio (–999 in Normal mode, Ratio: dB or % depending on Ratio Mode setting)
NOTE
Chapter 4
MRECEIVE
Measure
:CONFigure:FCOunter
This key invokes Frequency Counter Measurement
:CONF:FCO
If you are using the ASCII data output format, you will not get the same number of
significant digits in the returned value as shown on the display. You must select one
of the binary block data formats. (See FORMat:TRACe:DATA)
Front-Panel Key and SCPI Command Reference
Measure
157

Advertisement

Table of Contents
loading

This manual is also suitable for:

N5531sE4440aE4443aE4445aE4446aE4447a ... Show all

Table of Contents