4.1.3 Using Applications in PSA Series vs. VSA E4406A

NOTE
This information only applies to the application modes:
Basic, cdmaOne, cdma2000, 1xEV-DO, W-CDMA, GSM, EDGE,
NADC, and PDC.
Command
*RST
Resets instrument, putting
it in continuous
measurement mode. Use
INIT:CONT OFF to select
single measurement mode
and INIT:IMM to start one
measurement.
CONFigure:
Accesses the measurement
<measurement>
and sets the instrument
settings to the defaults. If
you were already in single
measurement mode, it
takes one measurement
and then waits.
*ESE default
Default is 255 which
means that every
error/status bit change that
has occurred will be
returned with a *ESR?
query. You must set the
value of *ESE to choose
only the bits/status that
you want returned.
*LRN
The command is not
available.
TRIGger
In Spectrum Analysis
commands
mode only one value can
be set for the trigger's
source, delay, level, or
polarity.
Basic, GSM, EDGE,
cdmaOne, cdma2000,
W-CDMA, NADC, PDC
modes function the same
as VSA
Chapter 4
Programming Command Compatibility Across Model Numbers and Across Modes
PSA Series
VSA E4406A: A.04.00
Resets instrument, putting it
in single measurement
mode. One measurement is
initiated when the command
is sent.
Same as PSA.
Accesses the measurement
and sets the instrument
settings to the defaults. If
you were already in single
measurement mode, it takes
one measurement and then
waits.
Default is 0 which means
that none of the error/status
bit changes that have
occurred will be returned
with a *ESR? query. You
must set the value of *ESE
to choose the bits/status that
you want returned.
The command is available.
You can select a unique
trigger source for each
mode. Each trigger source
can have unique settings for
the its delay, level, and
polarity.
Front-Panel Key and SCPI Command Reference
VSA E4406A: A.05.00
Resets instrument, putting it
in single measurement
mode. No measurement is
initiated when the
command is sent. Use
INIT:IMM to start one
measurement.
Accesses the measurement
and sets the instrument
settings to the defaults. If
you were already in single
measurement mode, it does
not initiate a measurement.
Use INIT:IMM to make one
measurement.
Same as VSA A.04.00.
Default is 0 which means
that none of the error/status
bit changes that have
occurred will be returned
with a *ESR? query. You
must set the value of *ESE
to choose the bits/status that
you want returned.
The command is available.
Same as VSA A.04.00.
You can select a unique
trigger source for each
mode. Each trigger source
can have unique settings for
the its delay, level, and
polarity.
123