Additional MyoBock components such as suction socket electrode, electrode cable, pull-in tube and tube valve for
suction socket as needed.

6 Preparing the product for use

6.1 Installing the AxonRotation Adapter

6.2 AxonMaster

Various control programmes and switching methods as well as the option of making refined patient-specific set­
tings are available for optimum adaptation to the patient. The standard programmes can be selected using the
560X500=* AxonSoft adjustment software and include predefined control options and switching methods.
Two MyoBock electrodes, e.g., 13E200/202=*, or if necessary a pull switch, e.g., 9X18, can be connected to the
13E500 AxonMaster. For information please refer to the 647G590 instructions for use for the 13E500 AxonMaster
Input on AxonMaster
Input 1
Input 2
Input 3
► Remove the cover caps from the required inputs. The remaining inputs cannot be used at this time and there­
fore remain covered by the caps.

6.2.1 Controlling the Electronic Lock

The electronic lock can be controlled myoelectrically by the following means configured at the factory:
A long co-contraction on channels 1 and 2 of the AxonMaster releases or locks the lock.
A short co-contraction switches the gripping mode of the gripper.
A pulse on channel 3 of the AxonMaster releases or locks the lock.
AxonArm Hybrid 12K500=*, AxonArm Ergo 12K501=*
1) Use silicone grease to seal the plug connection.
2) Connect the AxonRotation Adapter with the con­
necting cable.
3) Position the AxonRotation Adapter such that the
middle screw holes on the lamination ring can
be used.
INFORMATION: The other screw positions can
be used for correcting the alignment of the
AxonRotation Adapter after the lamination ring
has been glued in.
4) Screw the AxonRotation Adapter to the lamination
ring.
Electrode
Closing the hand
Opening the hand
Pull switch (optional)