5 Preparing the product for use

5.1 Information on fabrication of a prosthesis
CAUTION
Incorrect alignment, assembly or adjustment
Injury due to incorrectly installed or adjusted as well as damaged prosthetic components
Observe the alignment, assembly and adjustment instructions.
Fabricating a prosthesis that is optimally tailored to the patient demands technical knowledge of
orthopaedic technology. The alignment and adjustments may only be carried out by a prosthetist.
The alignment and adjustments are mainly completed in the following steps:
1. Bench alignment
2. Static alignment
3. Dynamic trial fitting
These steps are first carried out with a test prosthesis in order to determine the best possible
combination and positioning of the prosthetic components relative to each other. Once all set­
tings have been tailored to the patient, the definitive prosthesis is produced. The same steps are
repeated for this purpose.
In dynamic trial fitting, the patient must be given sufficient time to become familiar with the func­
tions of the prosthesis, thereby learning safe use for everyday life.
5.2 Information on Using a Cosmetic Cover
CAUTION
Use of talcum
Risk of injury, damage to the product due to lack of lubrication
Do not use talcum on the product or other prosthesis components.
► Only use the foam cover approved by the manufacturer (see section "Combination Possibilit­
ies" – see page 7).
► To reduce friction and to eliminate noise, apply 519L5 Silicone Spray directly onto the contact
surfaces of the cosmetic foam cover.
► After fabricating the cosmetic cover, repeat the dynamic trial fitting since the cover affects the
adjustments of the prosthesis (e.g. damping).
5.3 Bench Alignment
+ = anterior placement / – = posterior placement (in relation to the alignment reference line)
Item
see fig. 5
Required equipment:
Alignment apparatus (e.g., 743L200 L.A.S.A.R. Assembly or 743A200 PROS.A. Assembly),
743A80 50:50 Gauge, 743S12 Heel Height Measuring Apparatus
Position the prosthetic foot in the alignment apparatus in accordance with the following values:
   
Heel height:
effective heel height (x) + 5 mm
10
Quant­
Designation
ity
Prosthetic Knee Joint
1
Pyramid Dome Protector Kit
Bench alignment process
Reference number
3R78
2Z11=KIT