Preparing the product for use

6.2.2 Connecting the Electrodes

6.2.3 Installing the AxonMaster in the Upper Arm Socket

The AxonMaster can be secured with adhesive tape or a hook-and-loop fastening strap. The AxonMaster can be
covered with film to provide additional protection against moisture ingress.
Should the available space in the upper arm socket be too small an area for the AxonMaster (lamination dummy)
must be reserved during lamination of the inner socket.

6.2.4 Sealing the prosthesis

1) After laminating the lamination ring, remove the grease and any dirt.
2) Lubricate the groove, inner edge and thread of the lamination ring with 633F30 Special Grease.

6.2.5 Connecting the AxonMaster to the EasyPlug

6.3 Installing the EasyPlug

28
1) Remove the cover caps from the required inputs.
2) Seal the slots on the AxonMaster with silicone
grease.
3) Connect the electrode cables to the AxonMaster.
Observe the slot assignments:
Slot 1: close
Slot 2: open
Slot 3: pull switch (optional)
1) Grease the plug of the AxonMaster cable with
633F11 silicone grease.
2) Connect the cable.
3) Apply 633F11 silicone grease around the plug and
blanking covers to seal.
1) Lubricate the groove, inner edge and thread of the
lamination ring with 633F30 special grease.
2) Push the EasyPlug with the attached cable into the
lubricated lamination ring and press in firmly. The
EasyPlug enclosure must engage completely.
→ You must be able to hear a loud click!