11 Appendices

11.1 Symbols Used

Declaration of conformity according to the applicable European directives
In some jurisdictions it is not permissible to dispose of these products with unsorted household
waste. Disposal that is not in accordance with the regulations of your country may have a detri­
mental impact on health and the environment. Please observe the instructions of your national
authority pertaining to return and collection.
Serial number (YYYY WW NNN)
YYYY – year of manufacture
WW – week of manufacture
NNN – sequential number
Medical device
Manufacturer

11.2 Fault Display and Fault Elimination

11.2.1 LED flashes red and beeper sounds for 3 seconds

Probable cause:
Error in the Axon-Bus prosthetic system.
Remedy:
1. Switch the Axon-Bus prosthetic system off and back on.
2. Check the plug connections in the socket.
3. Connect the charger. If the charger displays the service symbol (yellow spanner) or the
11.2.2 The LED flashes three times after pressing the button in the colour of the current battery
capacity
Probable cause:
The connection from the AxonEnergy Integral to the Axon-Master has been interrupted.
Remedy:
1. Check the connection (e.g. plug connections, flex cable, EasyPlug etc.).
2. If no fault can be found, send the Axon-Bus prosthetic system to Ottobock Myo-Service.
11.2.3 LED alternately flashes yellow/red and beeper sounds for 3 seconds
Probable cause:
The Axon-Bus prosthetic system is overheated.
Remedy:
1. Allow the Axon-Bus prosthetic system to cool for a few minutes.
2. Do not use the Axon-Bus prosthetic system at high ambient temperatures.
11.2.4 The movements of the Axon-Bus gripping component are becoming increasingly slower
The patient is alerted to the diminishing battery charge level as the Axon-Bus gripping component increasingly
slows down and the gripping force decreases. Finally, when there is very little battery capacity remaining, the
Axon-Bus prosthetic system switches off automatically to protect the battery against harmful deep discharge.
11.3 Axon
The term "Axon" stands for Adaptive eXchange Of Neuroplacement data. The Axon-Bus is an Ottobock innovation
for the field of exoprosthetics: a data transmission system, derived from safety-related bus systems in the aviation
and automobile industries. For the user this means enhanced safety and reliability because of a considerably
reduced sensitivity to electromagnetic interference in comparison with conventional systems.
AxonArm Hybrid 12K500=*, AxonArm Ergo 12K501=*
system error symbol (red spanner) send the Axon-Bus prosthetic system to Ottobock Myo-
Service.