To Check The Power-Up Tests - Agilent Technologies 1670G Series User Manual

Logic analyzers
Hide thumbs Also See for 1670G Series:
Table of Contents

Advertisement

To check the power-up tests

The logic analyzer automatically performs power-up tests when you
apply power to the instrument. The revision number of the operating
system shows in the upper-right corner of the screen during these
power-up tests. As each test completes, either "passed" or "failed"
prints on the screen in front of the name of each test.
1 Disconnect all inputs, then insert a formatted disk into the
flexible disk drive.
2 Let the analyzer warm up for a few minutes, then cycle power by
turning off then turning on the power switch.
If the analyzer is not warmed up, the power-up test screen will
complete before you can view the screen.
3 As the tests complete, check if they pass or fail.
The Flexible Disk Test reports No Disk if a disk is not in the disk drive.
Performing Power-Up Self-Tests
passed
passed
passed
passed
passed
passed
No Disk
ROM text
RAM test
Interrupt test
Display test
PS2 Controller Test
Hard Disk Test
Flexible Disk Test
Operator's Service
Troubleshooting
461

Advertisement

Table of Contents
loading

Table of Contents