4.3 Calibration Wizard

For the user's convenience a TTC/12TC-wizard is included in the software.
To start the wizard, please use the magic wand from the Main Window's
toolbar or select Extras
selection dialog appears and lets the user choose between:
1.
Differential TTC (DTTC, available with differential measurement modes)
2.
Single-ended TTC (for channel 1 only, use DTTC for channel 2)
3.
Save current TTC/12TC for immediate reload
4.
Reload last known TTC/12TC
5.
Reset currently loaded TTC/12TC
6.
Full 2-port calibration (only if F2P measurement module is available)
If you choose 1., 2., or
sary steps towards a reliable TTC, DTTC or 12TC; nevertheless subsequently
some helpful remarks are enlisted:
In the start window, an existing TTC/12TC can be reloaded. Please
ensure that the currently used measurement setup (in the first place
the attached cables and time base settings of the device) is identical
to that of the original TTC/12TC.
If a TTC/12TC is loaded from a previously wizard-saved file, the wiz-
ard jumps directly to the last window and you are done. This is only
necessary if different TTCs/12TCs (e.g. for different time bases) shall
be handled in parallel.
A TTC/12TC can be reloaded from wizard-saved TTC mas files and also
be imported from standard project mas files.
Three-Term Correction from the main menu. A
6.
the TTC-wizard will guide you through the neces-
T3SP-Series Operator's Manual
39
931776 RevE