Analog Output Signal Type And Range Selection; Table 6-9: Analog Output Current Loop Range; Table 6-10: Analog Output Pin Assignments - Teledyne 200E Instruction Manual

Chemiluminescence nitrogen oxides analyzer
Table of Contents

Advertisement

Model 200E Instruction Manual
The following DC current output limits apply to the current loop modules:
Table 6-9:
Range
0-20 mA
These are the physical limits of the current loop modules, typical applications use 2-20 or 4-20
mA for the lower and upper limits. Please specify desired range when ordering this option. The
default offset for all ranges is 0 mA.
Pin assignments for the ANALOG output connector at the rear panel of the instrument:
See Figure 6-1 for a visual representation and location of the analog output connector.

6.7.3.1. Analog Output Signal Type and Range Selection

To select an output signal type (DC Voltage or current) and level for one output channel
press SETUP - MORE - DIAG - ENTR - NEXT - NEXT and then:
PREV
These keys set
DIAG AIO OUTPUT RANGE:
the signal level
and type of the
selected channel
044100102 Rev A
Analog Output Current Loop Range

Table 6-10: Analog Output Pin Assignments

Pin
Analog
output
1
A1
2
3
A2
4
5
A3
6
7
A4
8
DIAG
ANALOG I / O CONFIGURATION
NEXT
DIAG AIO
AOUTS CALIBRATED: NO
< SET
CAL
SET>
DIAG AIO
CONC_OUT_2:5V, CAL
< SET SET>
EDIT
DIAG AIO
CONC_OUT_2 RANGE: 5V
SET>
EDIT
5V
0.1V
1V
5V
10V CURR
Minimum Output
0 mA
VOLTAGE
CURRENT
Signal
Signal
V Out
I Out +
Ground
I Out -
V Out
I Out +
Ground
I Out -
V Out
I Out +
Ground
I Out -
V Out
not available
Ground
not available
ENTR
EXIT
EXIT
EXIT
EXIT
ENTR
EXIT
Operating Instructions
Maximum Output
20 mA
Press SET> to select the
analog output channel to be
configured. Press EDIT to
continue
Pressing ENTR records the new setting
and returns to the previous menu.
Pressing EXIT ignores the new setting and
returns to the previous menu.
61