Table 6-6: Analog Output Voltage Range Min/Max; Table 6-7: Analog Output Current Loop Min/Max; Table 6-8: Analog Output Pin Assignments - Teledyne 300E Instruction Manual

Carbon monoxide analyzer
Table of Contents

Advertisement

Model 300E Carbon Monoxide Analyzer
Instruction Manual
Range Span
The following DC Current output options are also available from the factory:
Range Span
Pin-outs for the Analog Output connector at the rear panel of the instrument are:
PIN
1
2
3
4
5
6
7
8
P/N 04288 REV: A
Table 6–6: Analog Output Voltage Range Min/Max
Minimum Output
0-100 mVDC
0-1 VDC
0-5 VDC
0-10 VDC
The default offset for all ranges is 0 VDC.
Table 6–7: Analog Output Current Loop Min/Max
Minimum Output
2-20 mADC
4-20 mADC
The default offset for all ranges is 0 mADC.
Table 6–8: Analog Output Pin Assignments
Analog Output
A1
A2
A3
A4 (Spare)
Maximum Output
-5 mVDC
-0.05 VDC
-0.25 VDC
-0.5 VDC
Maximum Output
1 mADC
3 mADC
VDC Signal
V Out
Ground
V Out
Ground
V Out
Ground
V Out
Ground
Operating Instructions
105 mVDC
1.05 VDC
5.25 VDC
10.5 VDC
21 mADC
21 mADC
mADC Signal
I Out +
I Out -
I Out +
I Out -
I Out +
I Out -
Not Available
Not Available