Table 6-12: Span Voltages And Adjustment Tolerances For Analog Output Signal Calibration - Teledyne 400E Instruction Manual

Ozone analyzer
Table of Contents

Advertisement

Model 400E Ozone Analyzer
Instruction Manual
To calibrate the outputs as a group press the following:
Exit at Any
Time to Return
to the main
DIAG Menu
If any of the channels have
not been calibrated this
message will read NO .
6.7.3.5. Manually Calibrating Analog Output Signal Levels
The analog outputs in voltage mode can be manually calibrated to closely match
the characteristics of the data recorder. Outputs configured for 0.1V full scale
should ALWAYS be calibrated manually.
Calibration is done through the instrument software in conjunction with a voltmeter
connected across the output terminals (see Figure 6-2). Adjustments are made
using the front panel keys. First the zero -point is set then the span-point
(Table 6-12).
The software allows this adjustment to be made in 100, 10 or 1 count increments.
Table 6-12: Span Voltages and Adjustment Tolerances for Analog Output
Full Scale
0.1 VDC
1 VDC
5 VDC
10 VDC
P/N 04316 Rev: B
STARTING FROM DIAGNOSTIC MENU
(see Section 6.6)
ANALOG I / O CONFIGURATION
DIAG
PREV
NEXT
DIAG AIO
AOUTS CALIBRATED:
< SET SET>
CAL
DIAG AIO
AUTO CALIBRATING CONC_OUT_1
AUTO CALIBRATING CONC_OUT_2
AUTO CALIBRATING TEST_OUTPUT
DIAG AIO
AOUTS CALIBRATED:
< SET SET>
CAL
Signal Calibration
Adjust Zero Within
±0.0005V
±0.001V
±0.002V
±0.004V
ENTR
EXIT
NO
EXIT
YES
EXIT
Span Voltage
Adjust Span Within
90 mV
900 mV
4500 mV
4500 mV
Operating Instructions
If AutoCal has been turned off for
any channel (see Section 6.7.3.5),
the message for that channel will
be similar to:
NOT AUTO CAL CONC_OUT_1
±0.001V
±0.001V
±0.003V
±0.006V
Exit to Return
to the I/O
Configuration
Menu
103