Summary Report

The Summary Report provides a condensed listing of selected data results. In the Selected Reports
list box, highlight Summary, then click Edit. Select the data types to include in the Summary report.
In the Pore Volume group box, if Adsorption total or Desorption total is selected, the p/p
enabled. Enter the relative pressure used to calculate the total pore volume.
If Use calculation assignments is not selected on the Collected Data window, the isotherm is interpo-
lated to the value in the p/p
Otherwise, the point selected with calculation assignment is used. Pass/Fail criteria can be specified
for up to four parameters
Field or Button
Select All / Deselect
All buttons
Item [n]
246-42800-01 - Aug 2013
fields and the point is used for the total pore volume calculation.
o
.
Selects (or deselects) all options.
Use to enable the first Pass/Fail item. Until the Summary Report is
selected, S A Single-point BET will be displayed by default. When the
checkbox is selected, click the Pass/Fail button and select criteria options
for pass/fail options.
• S A: Single-point BET checkbox - use to enable the Pass/Fail [n]
button in the Item [n] group box.