ASAP 2460 Operator's Manual
Action:
6230-
Dosing method does not match analysis type.
Cause:
Action:
6231-
Dosing method does not match analysis type.
Cause:
Action:
6232-
Po gas in sample file does not match Po gas in other active samples.
Cause:
Action A:
Action B:
246-42800-01 - Dec 2013
Select another sample information file.
You tried to perform a nitrogen or nitrogen-like analysis with From
last port selected as the dosing method.
Change the Dosing Method to Normal in the Adsorptive Properties
dialog, then start the analysis.
You tried to perform a Krypton analysis with Normal selected as the
dosing method.
Choose From last port for Dosing Method in the Adsorptive Proper-
ties dialog, then start the analysis.
You tried to start an analysis using a sample file that contains a Po gas
other than the Po gas currently being used. All current analyses must
use the same Po gas.
Change the gas in the sample file to use the current Po gas.
Wait until the current analyses are complete, then run the sample.
Appendix B
B-37