Appendix B
6223-
There is no Nitrogen attached to the unit.
Cause:
Action:
6226-
An analysis cannot be started because all ports are currently is use.
Cause:
Action:
6227-
File (name) cannot be analyzed. It is currently being edited.
Cause:
Action:
6228-
Analysis gas in sample file does not match analysis gas in other active
samples.
Cause:
Action A:
Action B:
6229-
This file already selected for the analysis.
Cause:
B-36
A calibration requiring nitrogen was attempted but the software does
not recognize that nitrogen is attached.
Ensure that a nitrogen gas bottle is installed at one of the analysis ports,
then Select Unit > Unit configuration and enter N2 for the appropriate
valve. Refer to
Configuration, page
Unit
You tried to start an analysis while all ports are in use.
Wait until the analysis completes on a port, then another analysis can
be started.
You tried to start an analysis using a sample information file that is cur-
rently being edited.
Close the edit session on the sample information file, then the file can
be used for analysis
.
You tried to start an analysis using a sample file that contains an anal-
ysis gas other than the gas currently being used. All current analyses
must use the same analysis gas.
Change the gas in the sample file to use the current analysis gas.
Wait until the current analyses are complete, then run the sample.
You selected a sample file that is already being used for an analysis.
ASAP 2460 Operator's Manual
4-29
.
246-42800-01 - Dec 2013