Appendix B
4057-
At least two data points are needed for Freundlich calculations.
Cause:
Action:
4058-
At least two data points are needed for Temkin calculations.
Cause:
Action:
Cause:
Action:
4059-
Fewer than 2 points available for MP-Method calculations.
Cause:
Action:
4060-
Sample (file name) contains no data points.
Action:
Cause:
B-28
Less than two data points have been selected for the Freundlich report;
at least two are required.
Select Freundlich points on the Collected Data dialog. If calculation
assignments are not being used, edit the Freundlich Report options,
Absolute pressure range in the sample file.
Less than two data points have been selected for the Temkin report; at
least two are required.
Select Temkin points on the Collected Data dialog. If calculation
assignments are not being used, edit the Temkin Report options, Abso-
lute pressure range in the sample file.
You have selected at least one point with a negative pressure value to
include in the Temkin report.
Negative pressure points void Temkin calculations. Open the sample
file and click the Collected Data tab. Locate the negative value(s) in
the Temkin column and deselect it (them).
At least two points are required for the MP-Method calculations.
Edit the calculation assignments for the MP-Method report.
An attempt was made to save a sample without collected data as a t-
curve or alpha-S curve.
Repeat the save as t-curve or save as alpha-S operation after opening
a sample that has collected data.
ASAP 2460 Operator's Manual
246-42800-01 - Dec 2013