4061-
The t-curve must contain at least 2 points.
Cause:
Action:
4062-
Error during report preparation.
Cause:
Action A:
Action B:
4063-
No subreports selected.
Cause:
Action A:
4067-
No data points are within the range of pressures in the reference isotherm.
Cause:
Action:
4068-
No points were selected for the f-Ratio report.
Cause:
Action:
246-42800-01 - Dec 2013
At least two points are required in a thickness curve definition.
Edit the thickness curve.
An internal processing error has occurred.
Contact your Micromeritics service representative.
In Report Options, edit the report and choose the desired subreport.
A report was selected, but none of the subreports for that report were
selected, meaning nothing will be done for that report.
If the report was selected in error, remove the selection.
There are no collected data points within the range of pressures in the
reference isotherm.
Select data points that are in the range of the reference isotherm, or
select a more appropriate reference isotherm.
The collected data column for the f-Ratio report does not have any
points selected.
Edit the collected data dialog and select points for the f-Ratio report.