Micromeritics ASAP 2460 Operator's Manual page 312

Accelerated surface area and porosimetry system
Table of Contents

Advertisement

Appendix B
4077-
Cannot get surface area for: (file name)
Cause:
Action A:
Action B:
4078-
Slope and Y-Intercept cannot be determined from the selected points.
Cause:
Action:
4135-
HOA file [filename] does not exist.
Cause:
Action:
4136-
HOA [filename] is corrupt.
Cause:
Action:
B-32
The Isotherm report for the named overlay file has Per gram selected
for the Volume Adsorbed, and the Isotherm report for the primary file
has a surface area option selected for the Volume Adsorbed.
Edit the Isotherm report for the named overlay file and select a surface
area option for Volume Adsorbed.
Click
on the Report options dialog of the primary file and
Overlays
remove the named overlay file from the list.
The Langmuir report cannot be generated from the selected points.
Edit the calculation assignments in the Langmuir column on the
Collected Data dialog.
The sample file in the Heat of Adsorption report list does not exist.
Go to Report > Heat of Adsorption. Click Add Samples, then select
the sample file.
The sample file in the Heat of Adsorption report is corrupt.
Go to Report > Heat of Adsorption. Select the correct sample file,
then click Remove Sample. Rerun the Heat of Adsorption report.