Checking and Cleaning the Analysis Port Dewars

When handling dewars, be sure to observe the dewar precautions outlined in
Dewars, page
Filling
Ice and suspended frost particles may accumulate in the bottom of an analysis port dewar. Particles or
deposits exceeding 1/4 in. (0.64 cm) in depth may jam between the bottom of the sample tube and the
bottom of the dewar, causing the dewar not to raise fully. Accumulations of fine particles impede liq-
uid nitrogen circulation around the bottom of the sample tube. This causes the sample temperature to
be slightly higher which, in turn, can cause pore volume measurement errors in those samples exhibit-
ing high isotherm slope above 0.97 relative pressure.
Accumulated ice is likely to melt and form a pool of water in the dewar if all liquid nitrogen evapo-
rates. The water must be removed; otherwise it will solidify when liquid nitrogen is added and could
press on the bottom of the sample tube causing breakage.
To ensure problems do not develop due to ice accumulation, check the dewar after each use. Clean
dewars on a weekly basis as follows:
1.
Remove the dewars and pour the liquid nitrogen from the dewars into an appropriate cryogenic
container.
Do not pour liquid nitrogen directly into a sink. Doing so may cause drain pipes to
burst.
2.
Rinse the dewars with warm water to melt any ice accumulation which may remain in the dewars
then dry thoroughly.

Replacing Analysis Port Filters

A porous metal filter is located in each analysis port. Using a contaminated filter on an analysis port
may extend the time required to achieve a vacuum at that port. More importantly, the contaminant may
adsorb or desorb during analysis, affecting the analysis results. A contaminated filter on an analysis
port may cause a leak test to fail (if the contaminant outgasses) or cause a free space reading to be
much lower than normal.
Before removing and installing a port filter, ensure that the port valve is closed.
Observe the analysis system schematic to verify valve status.