Micromeritics ASAP 2460 Operator's Manual page 122

Accelerated surface area and porosimetry system
Table of Contents

Advertisement

Analysis Conditions Files
Field or Button
Equilibration button
3-28
Click to enter values for relative or absolute pressure.
Column heading
reads Absolute
Pressure (mmHg)
when Absolute
Pressure Dosing is
selected
• Relative Pressure (p/p
the equilibration interval will be applied.
• Equilibration Interval - the number of seconds between successive
pressure readings during equilibration. At least 11 pressure readings for
each equilibration are required to obtain an accurate reading of the
pressure changes.
Long equilibration intervals tend to lengthen analyses, but improve data
integrity.
Short equilibration intervals produce a faster analysis but may reduce
the accuracy of data.
• Minimum equilibration delay at p/p
of seconds required before equilibration can occur for a relative pressure
greater than or equal to 0.995. This field is not available if Absolute
pressure dosing is selected on the Analysis Conditions tab.
ASAP 2460 Operator's Manual
Description (continued)
) or Absolute Pressure (mmHg) - the pressure
o
> =0.995 - the minimum number
o
Does not
display when
Absolute
Pressure
Dosing is
selected.