Micromeritics ASAP 2460 Operator's Manual page 306

Accelerated surface area and porosimetry system
Table of Contents

Advertisement

Appendix B
4045-
Points in the report pressure table lie outside the collected data.
Cause:
Action:
4046- (file name) could not be opened for reading.
Cause:
Action:
4047-
Warning: An error occurred while reading (file name).
Cause:
Action:
4050-
None of the requested quantities adsorbed is within the range of the collected
data of more than one sample file.
Cause:
Action:
4051-
The sample (file name) does not have any data in the range of the requested
quantities adsorbed.
Cause:
Action:
B-26
Calculation assignments are not being used and more than one of the
report pressure table points is above the range of the collected data, and
more than one is below.
Change the report pressure table to be more consistent with the col-
lected data.
A thickness curve file could not be opened.
If the problem persists, restart your computer and optionally perform a
media integrity check (using ScanDisk).
An error happened during a read operation of a thickness curve file.
If the problem persists, restart your computer and optionally perform a
media integrity check (using ScanDisk).
The files selected for HOA reporting cannot produce a valid report.
Choose a different set of files.
The sample's data cannot be interpolated to any of the quantities
adsorbed.
Edit the quantities adsorbed so that they are within the range of the col-
lected data.
ASAP 2460 Operator's Manual