Micromeritics ASAP 2460 Operator's Manual page 253

Accelerated surface area and porosimetry system
Table of Contents

Advertisement

ASAP 2460 Operator's Manual
3.
Drain the used oil:
a.) Place a waste container under the drain spout.
b.) Remove the plug from the drain spout.
c.) Allow the oil to drain into the waste container.
d.) Replace the drain plug.
4.
Remove the oil-fill plug.
Adding fluid above the midway position on the fluid level indicator may cause fluid
to splash into the vacuum hoses and leak from the vacuum pump.
5.
Add fresh fluid to the oil-fill port until the level is midway between the two indicator lines in the
oil-level window.
6.
Check the washer or O-ring used at the oil-filling port; replace if necessary.
246-42800-01 - Aug 2013
Preventive Maintenance
Oil-fill level
Drain plug
Oil-fill port
7-11