Micromeritics ASAP 2460 Operator's Manual page 300

Accelerated surface area and porosimetry system
Table of Contents

Advertisement

Appendix B
4022-
The entered bath temperature value (Po and Temperature Options of the Anal-
ysis Conditions) is outside the range of the temperatures listed in the Psat vs.
Temperature Table (Adsorptive Properties).
Cause:
Action A:
Action B:
4023-
The file (sample file) cannot be prepared for analysis. It is open for editing and
contains errors.
Cause:
Action:
4024-
Backfill gas in sample file does not match any gas in unit.
Cause:
Action:
4025-
There is no Helium attached to the unit.
Cause:
Action:
B-20
The Po that corresponds to the entered temperature cannot be found
because the entered temperature is outside of the range of values in the
Psat vs. T table.
Enter a temperature that is within the range of the table.
Use a different Psat vs. T table.
You attempted an analysis using a sample file that is being edited and
has invalid values in one or more of its fields.
Enter valid values into the fields that have errors, save the changes, and
close the dialog box.
You selected a backfill gas that is not included in the gas configuration.
Make sure that gas selections match the gases that are attached to the
instrument. Choose one of the gases as the backfill gas, or edit the sam-
ple file and choose one of the existing gases as the backfill gas.
A measured free space has been requested but helium does not appear
to be available.
Make sure the helium cylinder is attached to the unit at its designated
port.