Micromeritics ASAP 2460 Operator's Manual page 285

Accelerated surface area and porosimetry system
Table of Contents

Advertisement

ASAP 2460 Operator's Manual
2471- Unit n - S/N; nn has an invalid communications port specified. It cannot be
initialized.
Cause:
Action:
2474- (Unit {number}) Communications port COM (port number) specified in the pro-
gram control files is already in use. Unit cannot initialize.
Cause:
Action:
2475- (Unit {number}) Communications port COM (port number) specified in the pro-
gram control files cannot be accessed. Unit cannot initialize.
Cause:
Action:
2476- Unit startup failed. Cannot initialize the communications port.
Cause:
Action:
2477-
(Unit n; Serial nn) did not properly initialize.
Cause:
246-42800-01 - Dec 2013
The communications port specified for this analyzer during installation
(or configuration) is invalid.
Use the setup program to change the analyzer configuration.
The communications port assigned to the indicated unit is in use by
another program.
End the program using the busy port or use the Setup program to
change the communications port assigned to this program.
The operating system has prevented the application from accessing the
communications port.
Review the hardware configuration of the computer, ensure that no
other application is using the port. Contact your service representative
if you continue to receive this error message.
The communications port specified during installation is invalid.
Use the setup program to change the analyzer configuration. Contact
your service representative if you continue to receive this error mes-
sage.
The software was unable to initialize the analyzer.
Appendix B
B-5