ASAP 2460 Operator's Manual
Ensure adequate ventilation when using solvents for cleaning purposes.
6.
Screw the end fitting back onto the trap and tighten securely by hand. Insert the O-ring and screw
the end fitting back onto the trap and tighten securely by hand.
7.
Make sure the centering ring is in place on the intake port.
8.
Place the trap on the centering ring.
Oil vapor trap
Centering ring
Intake port
9.
Open the clamp and place it around the flange of the intake port and the flange of the trap. Swing
the clamp fastening screw toward the intake port until it fits into the slot in the other half of the
clamp. Tighten the wing nut securely by hand.