The analysis gas (name) does not match the model gas (name).
Cause:
Action:
4074-
The analysis temperature (value) does not match the model temperature
(value).
Cause:
Action:
4075-
The models cannot be located in the models folder. Reinstall the software.
Cause:
Action:
4076-
Invalid sample file - Adsorptive Properties "Non-condensing adsorptive" can-
not be used unless Analysis Conditions "Absolute pressure dosing" is selected.
Cause:
Action A:
Action B:
246-42800-01 - Dec 2013
The model assumes a specific gas and the sample file uses a different
one.
Select a model that assumes the same gas.
The wrong model was selected for DFT reporting.
Select a model for which the analysis temperature matches.
There are no models in the MODELS subdirectory (of the program
directory). These are placed in the directory during software installa-
tion.
Reinstall the software to place the DFT models in the MOELS direc-
tory.
You selected a sample file that has Non-condensing adsorptive
selected in the Adsorptive Properties but does not have Absolute pres-
sure dosing selected in Analysis Conditions.
If you plan to use a non-condensing adsorptive, select Absolute pres-
sure dosing on the Analysis Conditions dialog.
If you are not using a non-condensing adsorptive, deselect the Non-
condensing adsorptive option on the Adsorptive Properties dialog, or
choose an adsorptive that you are sure in not non-condensing.