Appendix B
4012-
Psat gas in sample file does not match Psat gas in unit.
Cause:
Action:
4014-
File (name) is not a valid file for conversion.
Cause:
Action:
4015-
Error creating export file for sample (filename).
Cause:
Action:
4016-
Sample (file name) has no data for export.
Cause:
Action:
B-18
You attempted to start an analysis using a sample information file in
which the Psat gas does not match the Psat gas entered in the unit con-
figuration.
If necessary, attach the appropriate gas bottle, then enter the gas in the
Unit Configuration dialog box.
The file selected for conversion is not a MICMOS 2400 or 2405 file.
Select only files that have been created by MICMOS 2400 or 2405
programs.
A file error occurred during creation of an export output file.
The output file name may be invalid. Ensure that the target directory
exists. Ensure that the target diskette or CD is not full or write pro-
tected. The target disk drive may be damaged or inoperative. Verify
that other files may be created on the same drive. Contact your service
representative if you continue to receive this error message.
The file selected for export has a status of No Analysis. No export file
will be created.
Select a file which contains analysis data.