ASAP 2460 Operator's Manual
Action:
Action:
2558-
The instrument is busy.
Cause:
Action:
Action:
2559-
SPC directory (name) does not exist.
Cause:
Action:
Action:
2560-
File (filename) cannot be created or opened. It has an unrecognized extension.
Cause:
Action:
246-42800-01 - Dec 2013
If the named directory was accidentally renamed or move, restore the
directory.
If the file reference is incorrect, remove the file reference.
The requested operation cannot be executed.
An instrument is busy. The requested operation cannot be executed.
The instrument is busy. The requested operation cannot be done at this
time.
Wait until the instrument is idle and try the operation again.
Files were selected for SPC reporting from a directory which no longer
exists.
If the directory was deleted, renamed, or moved in error, restore the
directory.
Select a different set of files for SPC reporting.
A file name with an extension unrecognizable by the software was
typed into a file selector.
Type in a different file name. The extensions used by the software are:
SMP - Sample File
ANC - Analysis Conditions
DEG - Degas Conditions
ADP - Adsorptive Properties
RPO - Report Options