Report Options Files
Field or Button
Overlays button
Import button
Apply thermal
transpiration
correction checkbox
Selected Reports list
box
Browse button
Cancel button
Close button
Edit button
OK button
Preview button
Save button
3-32
Refer to
Multiple Sample
Use to import up to 25 pore distribution data files. These datasets are
shown only in BJH and Dollimore-Heal reports.
Use to correct the temperature-induced pressure difference between the
manifold and the chilled sample tube. This option is most significant for
pressures less than approximately 1.0 mmHg. Do not use filler rods in the
sample tube when applying correction for thermal transpiration.
Always use thermal transpiration when performing micropore analyses.
Refer to CALCULATIONS, page
thermal transpiration.
• Inside diameter of sample tube text box - Enabled when Apply
thermal transpiration correction is selected. Enter the inside
diameter of the sample tube used in the analysis.
Select the checkbox to the left of the report names to include in the report.
For BJH reports, BJH pore dimension can be calculated in pore width (w),
pore radius (R) or pore diameter (D). Go to Options > Units to specify
default calculations.
Refer to
Common Fields and Buttons - File Menu
ASAP 2460 Operator's Manual
Description (continued)
Overlays, page 2-53.
for additional information on
C-1
Options, page 3-1.
246-42800-01 - Aug 2013