ASAP 2460 Operator's Manual
Field or Button
Surface area correction factor
text box
Fitted thickness range text boxes Enter the minimum and maximum thicknesses (in angstroms or
Tabular report checkbox
t-Plot checkbox
Pressures button
Cancel button
From / To text boxes
OK button
Table buttons
246-42800-01 - Aug 2013
Description (continued)
Enter the value to correct for surface areas that are not smooth.
This brings the values for BET surface area and micropore
surface area into accordance. For most samples, the default
value of 1.000 is adequate.
nanometers) to include in the thickness curve. Go to Options >
Units to specify default units. Refer to
Use to have a tabular report of data generated.
Use to have a graphical representation of data generated.
• Overlay samples checkbox - use to overlay sample files on
the t-plot.
• Autoscale x-axis - the x-axis field shows the statistical
thickness of the adsorbed film.
• Autoscale y-axis - the y-axis field shows the quantity of gas
adsorbed.
Refer to
button, page 3-30.
Pressures
Refer to
Common Fields and Buttons - File Menu
page 3-1.
Report Options Files
MENU, page 4-1.
UNIT
Options,
3-45
Need help?