SPX RADIODETECTION 6100-Cu Operation Manual page 93

Multifunction phone network analyser and copper tester
Hide thumbs Also See for RADIODETECTION 6100-Cu:
Table of Contents

Advertisement

action will terminate the ongoing measurement
test.
Frequency allows you to set the frequency
value for the Near End Crosstalk (NEXT).
Frequency Step lists a choice of frequency
step values.
Termination Resistance lists a choice of
resistance values for the termination of the
signal.
Show hookup diagram on test setup - If Yes,
the Confirmation page is prompted before
every NEXT test is started. Select No to hide
the diagram next time you run the test.
Hookup Diagram button opens a
Confirmation page displaying how to connect
your cables at the beginning of the test. When
FED is Enabled on the FED Setup page, both
the T-R/A-B and T1-R1/A1-B1 pairs will be
terminated at the FED. The hookup diagram
uses graphic and textual components.
o
Proceed button closes the hookup
diagram prompt and starts the test.
o
Cancel button closes the hookup
diagram prompt only.
o
The Confirmation window is prompted
before every NEXT test is started if you
selected Yes for the Show hookup
diagram. You can select No to hide the
diagram next time you run the test.
However, you can check the connection
diagram at any time by pressing the
Hookup Diagram button. In this case,
the test does not start when you press
the OK button to close the Confirmation
window.
© 2020 Radiodetection Ltd
6100-Cu Multifunction Copper Tester Operation Manual
Threshold Settings button opens a new page
that allows you to set the NEXT Spot Threshold
for the test.
Step -> Freq button automatically
enters/transfers the Frequency Step value into
the Frequency value. If the step size is less
than 10 kHz (but not 4.3125 kHz), the
Frequency value will be set to 10 kHz. If the
Frequency Step is set to 4.3125 kHz, the
Frequency value will be set to 8.6250 kHz.
Result displays the pass/fail status, NEXT and
Frequency (TX) values.
Test Status displays the current state of
measurement or an error message.
To select the Spot parameter values:
1. Press the up/down arrow keys to highlight the
desired parameter.
2. Press the left/right arrow keys to view and
select the options.
OR
3. Press
on a value to open a list box of
options or the alphanumeric editor screen and
use the navigation keys to scroll through.
4. Press
to confirm the value.
Spot Threshold Settings
The Threshold Settings page allows you to set the
threshold values for the NEXT Spot test.
Type is the threshold limit: either Maximum or
Minimum.
Enabled is a flag to use the current threshold:
either Yes or No.
93

Advertisement

Table of Contents
loading

Table of Contents