the Calculated Attn, plus half of the
Threshold Tolerance.
When Mask is enabled for the selected
bandwidth:
o
The end of tolerance or fail point is
displayed as a red cursor value.
o
Good is the good balance threshold
mask for the currently selected
bandwidth, represented by green in the
graph.
Test Status displays the current state of
measurement or an error message.
To select the parameter values:
1. Press the up/down arrow keys to highlight the
desired parameter.
2. Press the left/right arrow keys to view and
select the options.
OR
3. Press
on a value to open a list box of
options or the alphanumeric editor screen and
use the navigation keys to scroll through.
4. Press
to confirm the value.
WB Attenuation Threshold Settings
The Threshold Settings page allows you to set the
WB Attenuation threshold values for the test.
Attenuation sets up the good attenuation
threshold parameters.
Type is the threshold limit for each pair.
Selecting Minimum or Maximum acts as the
© 2020 Radiodetection Ltd
6100-Cu Multifunction Copper Tester Operation Manual
single point pass/fail threshold flat mask while
selecting Mask provides a shaped mask as
threshold. When Mask is selected, it will be
used as pass/fail criteria next time the test is
run.
Enabled is a flag to use the existing threshold:
either Yes or No. For example, setting the
threshold Enabled to No will disable pass/fail
threshold.
To set threshold values:
1. Press the up/down arrow keys to highlight the
desired parameter.
2. Press the left/right arrow keys to view and
select the options.
OR
3. Press
on a value to open a list box of
options or the alphanumeric editor screen and
use the navigation keys to scroll through.
4. Press
to confirm the value.
Cable Configuration
The Cable Configuration page is only required if using
theoretical calculated attenuation pass/fail. The page
allows you to set cable parameters for the WB
Attenuation test. Each time the Cable Configuration
is changed, the total equivalent attenuation table will be
recalculated and redrawn.
No. of Cable Sections allows you to set the
number of consecutive cable sections, to a
maximum of 4. If unknown, set the number to
1.
105