6100-Cu Multifunction Copper Tester Operation Manual
if you stop the test. The last valid results then remain
displayed.
Current Group is the current test group name,
for example Noise Tests.
Current Test displays the test being
measured. Here you can switch to another test
from the given list. However, this action will
terminate the ongoing measurement test.
VF Noise Filter allows you to define the voice
frequency noise filter to be used, dependent on
ITU or ANSI Standard mode and the Noise
Filter set in Setup/Application Settings.
Termination Resistance is a read-only value of
600 Ω.
Test Duration allows you to set the length of
time to run the test from a list of values. If
UFED Control option is enabled, the maximum
value is limited to15 min.
Threshold Settings button opens a new page
that allows you to set the VF impulse noise
thresholds.
Result displays the High, Mid, Low counts,
respective detection levels, and Elapsed Time.
The pass/fail status of the Mid level is
displayed after 1 s and then periodically
refreshed, until you stop the test.
Test Status displays the current state of
measurement or an error message.
86
To select the parameter values:
1. Press the up/down arrow keys to highlight the
desired parameter.
2. Press the left/right arrow keys to view and
select the options.
OR
3. Press
on a value to open a list box of
options.
4. Press
to confirm the value.
VF Impulse Noise Threshold Settings
The Threshold Settings page allows you to set the VF
Impulse Noise Threshold values for the test.
Low Counter Power Detection Level defines
the low threshold limits for the impulse noise
test.
Power Separation defines the level difference
between the Low, Mid, and High thresholds.
Specify a value between 1 and 6 dB.
Mid Counter allows you to set the threshold
value for this counter.
Type is the threshold limit of the Mid Counter:
either Maximum or Minimum.
Enabled is a flag to use the current Mid
Counter threshold: either Yes or No.
To set threshold values:
1. Press the up/down arrow keys to highlight the
desired parameter.
2. Press the left/right arrow keys to view and
select the options.
OR
© 2020 Radiodetection Ltd