9. Verify that the calibration weight is set to 50 and click on
FmXP will prompt you to place the weight on the scale.
10. Using gloves or tweezers, place the 50 g calibration weight on the pedestal.
11. Replace the scale cover and click
FmXP will display the following message (Figure 5-9):
12. Upon completion, FmXP displays the following message:
13. Remove the weight from the scale and click
5-8
.
OK
Figure 5-9 FmXP Calibrating Scale Message
Figure 5-10 FmXP Calibration Complete Message
.
Calibrate
.
OK
Calibration and Adjustment