Configuration - Agilent Technologies 6850 Series II User Information

Gas chromatograph
Hide thumbs Also See for 6850 Series II:
Table of Contents

Advertisement

Caution
Do not install a pressure regulator on the CO
would occur in the regulator instead of the oven.
Correct
Warning
All fittings and tubing should be stainless steel. The pressure inside a liquid
CO
2
1.
2.

Configuration

In order to properly control instrument gas flows and temperatures, and to
provide proper functionality for valves, the instrument maintains a list of
manually-configured settings. These include carrier and detector gas types,
valve types (if installed), and the time and date. Agilent sets these parameters
at the factory. In general, if using the same gases as used for the detector
checkout method, no further configuration is needed.
Released: March 2004
Dip tube
tank can be as high as 1000 psi.
Connect a CGA 320 to 1/4-inch male NPT fitting to the tank followed by a
Swagelok part no. SS-200-7-4 1/4-inch female NPT to 1/8-inch tube fitting.
Connect the tank to the cryogenic filter inlet with 1/8-inch diameter heavy-
wall stainless steel tubing. The tubing may be up to 50 feet long.
Allow some slack in the line to make movement of the tank or GC easier.
Coil and fasten the ends of the tubing to keep it from "whipping" if it
breaks.
6850 Series II GC User Information
tank, as vaporization and cooling
2
Pressure regulator
Incorrect
Installation
Configuration