Shutdown Messages - Agilent Technologies 6850 Series II User Information

Gas chromatograph
Hide thumbs Also See for 6850 Series II:
Table of Contents

Advertisement

WARNING 501
WARNING 502
WARNING 503

Shutdown messages

SHUTDOWN 1
Possible causes:
Released: March 2004
Change liner
The service limit (number of injections) for the liner has been exhausted. To
maintain optimum performance, use your control module to go to Status/
Service/ Start Service when ready to replace the liner.
The GC will still start a run when this warning is displayed.
To adjust/turn off the service limit, go to Status/Service/Service Limits.
Service syringe
The service limit (number of injections) for the syringe has been exhausted. To
maintain optimum performance, use your control module to go to Status/
Service/ Start Service when ready to service the syringe.
The GC will still start a run when this warning is displayed.
To adjust/turn off the service limit, go to Status/Service/Service Limits.
Service column
The service limit (number of injections) for the column has been exhausted. To
maintain optimum performance, use your control module to go to Status/
Service/ Start Service when ready to service the column.
The GC will still start a run when this warning is displayed.
To adjust/turn off the service limit, go to Status/Service/Service Limits.
Oven shut off
To clear the shutdown, turn the GC off and then on again or change the oven
temperature.
Check the oven flaps. If an oven flap is stuck open completely or partially, it is not
operating correctly. During normal use, the flaps should be full open or full closed.
Oven lid down but not latched
Defective main board
Defective heater
The power required to keep the oven at setpoint exceeds the expected power
for that temperature. The oven is shut off and the flaps should be opened
half-way for cooling.
6850 Series II GC User Information
Messages
Shutdown messages