Agilent Technologies 6850 Series II User Information page 85

Gas chromatograph
Hide thumbs Also See for 6850 Series II:
Table of Contents

Advertisement

WARNING 107
WARNING 109
WARNING 111
WARNING 113
WARNING 114
WARNING 115
Possible causes:
WARNING 116
WARNING 117
WARNING 119
WARNING 121
Possible cause:
WARNING 128
Possible cause:
WARNING 130
Possible cause:
Released: March 2004
Det config changed
Inlet config changed
Column config changed
Aux 3 config changed
Aux 4 config changed
Aux 5 config changed
There is a new detector/inlet/column/auxiliary EPC module in the GC
The detector/inlet/column/auxiliary EPC module in the GC does not match the method
The method was changed to default detector/inlet/column/auxiliary EPC
module values because either a different one was installed, or a newly loaded
method uses a different detector/inlet/column/auxiliary EPC module than is
available.
If a new detector/inlet/column/auxiliary EPC module, the GC will enter default
setpoints for the new type.
If the method uses a different detector/inlet/column/auxiliary EPC module, the
method is unchanged and the warning message flashes.
Run log full
The Run log can hold 50 entries of information. When it reaches the maximum,
this warning will appear.
Inlet calib deleted
Det calib deleted
Aux calib deleted
Power failure during recalibration
While entering a new flow module calibration, the GC crashed or encountered
a power failure before the recalibration was complete. The warning appears
when the instrument is turned on again.
The calibration for the module returned to the default. Recalibrate.
Inlet flow cal failed
Defective or leaking flow module
An attempted auto calibration of the flow sensor offset for the inlet exceeded
the allowable calibration range. The previous calibration setpoints remain in
memory.
Aux cryo disabled
Incorrectly installed inlet cryo valve or valve box
The inlet cryo valve is also being used by aux thermal zone 1. The valve is
disabled.
6850 Series II GC User Information
Messages
Warning messages
page 85 of
256