7.
Pulling up on the
knurled nut removes
entire top of inlet.
O-ring
8.
9.
10. Replace the top of the inlet and tighten the knurled nut. Replace the
11. Restore the GC to normal operating conditions.
Replacing the inlet flow module
Replacing the purged packed inlet flow module is similar to replacing a
split/splitless inlet flow module. See
page
Released: March 2004
Loosen the knurled nut and pull it upward. The O-ring will be visible.
Replace it if it is hard and brittle or cracked. See
page 142
for details.
Open the oven lid. Use a light source to illuminate the inside of the inlet
from inside the oven while looking through the inlet from the top. If
deposits are present, they should be visible.
Insert the brush into the inlet. Scrub the interior walls of the inlet
vigorously to remove all deposits. You may need to wet the brush with sol-
vent. Use a burst of compressed air or nitrogen to dry the inlet and remove
loose contaminants.
column. See
"Columns and Traps" on page 21
124.
6850 Series II GC User Information
Maintaining a purged packed inlet
"Changing the O-ring" on
for details.
"Replacing the inlet flow module" on
Routine Maintenance: Inlets
page 149 of
256