Align the turret
The alignment procedure makes sure that your 6850 ALS is ready for operation
using the installed turret.
Always perform an alignment after changing turret types (from the 2-mL vial
turret to the 4-mL vial turret, and vice-versa).
1.
2.
3.
4.
Released: March 2004
Install one mounting screw, then rotate the turret. The outer edge should
remain level. Install the remaining screws.
Reinstall the ESD cap.
Align the turret. See
Remove all vials from the turret. Open the injector door and remove the
syringe.
Remove the needle support assembly. See
assembly" on page
Push the align button. The injector and turret will perform a series of
checks.
Loosen the screws in the clamps that hold the calibration probe to the
bottom of the turret and remove the probe. Do not remove the clamps.
6850 Series II GC User Information
"Align the turret" on page
224.
Align button
Automatic Liquid Sampler and Injector
The 6850 automatic liquid sampler
234.
"Replacing the needle support
page 234 of
256