Installing columns in the purged packed inlet
Agilent Technologies recommends that you avoid using capillary PLOT col-
umns with a purged packed inlet with a gas sampling valve unless you plumb
the valve directly to the column, bypassing the inlet.
Before installing a column in this inlet, be sure you have a capillary liner and
glass insert installed. See
insert" on page
Otherwise, begin with Step 4.
Released: March 2004
Adjust the column position so that the mark on the column is even with
the bottom of the column nut.
Tighten the column nut an additional 1/4 to 1/2 turn beyond finger tight so
that the column cannot be pulled from the fitting with gentle pressure.
1/4 turn
138. If your insulation cup is not installed, begin with Step 1.
6850 Series II GC User Information
"Installing a liner" on page 137
Columns and Traps
Installing capillary columns
and