GC setup
Warning
If the unit has been in operation, areas may be hot enough to cause serious
burns. Turn off the heated zones and the oven and allow sufficient time for the
unit to cool.
1.
2.
3.
4.
ChemStation setup
1.
Path
View/Run Method/Control View/Instrument/
Edit Parameters/Options
View/Run Method/Control View/Instrument/
Edit Parameters/Signals
Released: March 2004
If you are using a split/splitless inlet, install a new septum, a new splitless
liner, and a new O-ring seal in the inlet. See
on page
120.
If you are using a purged packed inlet, install a new liner and glass insert
for the checkout column. Also install a new septum and O-ring. See
"Changing septa" on page 121
If you are using a cool on-column inlet, install a new septum, then check
for a proper fit of the needle in the column. See
page 121
and
"Cleaning the inlet" on page
If you are using a PTV inlet with the septum head, install a new septum or
microseal, inlet adapter, silver seal, and glass liner. See
septum" on page
156,
inlet adapter" on page
If you are using a PTV inlet with the septumless head, clean the head and
replace the Teflon ferrule and glass liner. See
head" on page
152,
"Replacing liners" on page
Install a 0.28-mm (0.011-in.) jet and/or the capillary column adapter, if
needed.
Install the 6850 evaluation column.
If present, remove the protective caps from the inlet manifold vents.
Bring up the proper online instrument for the 6850 to be tested. Make
these settings:
6850 Series II GC User Information
and
"Changing the O-ring" on page
"Replacing liners" on page
150.
"Replacing the Teflon ferrule" on page
157.
Setting
Keep instrument keyboard locked after method is
loaded? = No
Assign Signal 1 to the detector
Choose Save Data All
Data Rate = 20 Hz
Verifying Performance
"To install a liner and O-ring"
"Changing septa" on
131.
"Changing the
157, and
"To replace the
"Cleaning the septumless
154, and
FID checkout
142.
page 62 of