Plunger carrier loop
Bearing clip
Syringe latch
Syringe carriage
Figure 46.
Installing the needle support assembly
8.
9.
Caution
Be careful not to bend the needle during installation.
Caution
Do not operate the injector or sampler without a syringe in place because the
free swinging syringe latch may interfere with the motor.
10. Install the syringe. See
11. Move the slide up and down to make sure the needle is aligned with the
Released: March 2004
Guide
Syringe
carriage
tracks
Align the bearing on the needle support assembly with the plastic bearing
clip to the right of the syringe latch and push gently on the bearing until
the assembly snaps into place.
Place the injector or liquid sampler back on the GC or on a parking post.
needle guide in the support foot. The needle should slide smoothly in the
needle guide.
6850 Series II GC User Information
Syringe
carriage
tracks
Cross section
of slide alignment
Slide
Bearing
"Installing a syringe" on page 222
Automatic Liquid Sampler and Injector
Slide
for details.
page 227 of
Syringes
Slide
256