The Fault light is on
The Fault and the Run lights are on
Align Mode light is on
Released: March 2004
Probable cause
The injector door is open.
Suggested actions
1.
Ensure that the injector door is closed.
2.
If the fault light stays on, obtain Agilent service.
Probable causes
•
The injector is mounted incorrectly on the mounting post
•
Incorrect mounting post
Suggested actions
1.
Ensure that the injector is mounted properly. For more information,
see
"Installing a sampler or injector on the 6850 GC" on page
2.
Ensure the correct mounting post is installed.
3.
If the fault lights stay on, obtain Agilent service.
Probable causes
•
The turret is not properly installed
•
The type of turret was changed while the power was on
•
The system was not initialized
•
There is an injector memory error
Suggested action
1.
Verify the turret is properly installed. See
page
236.
2.
Perform the alignment procedure to initialize the system. See
the turret" on page
6850 Series II GC User Information
237.
Automatic Liquid Sampler and Injector
"Changing the turret" on
"Aligning
page 240 of
Faults
242.
256