Agilent Technologies 6850 Series II User Information page 143

Gas chromatograph
Hide thumbs Also See for 6850 Series II:
Table of Contents

Advertisement

Materials needed:
1.
2.
3.
Pulling up on the
knurled nut removes
entire top of inlet.
O-ring
4.
Released: March 2004
Gloves (if the inlet is hot)
A new Viton O-ring (part no. 5080-8898)
Septum nut wrench (part no. 19251-00100)
Forceps or tweezers (optional)
Load the SERVICE method or reset the inlet and oven temperatures to
OFF or turn the main power switch off. Let the heated zones cool to room
temperature. Also, make sure that:
If you have entered parameters that you do not want to lose, store
them as a method
If the detector is on, turn it off
Turn the inlet pressure off at the source.
If the inlet is hot, use the septum nut wrench. Loosen the knurled nut
completely. Pull up on the nut to remove the top portion of the inlet.
The O-ring will be visible. Remove the old O-ring. You may need to use
forceps to grab it. Using the tweezers, insert the new O-ring.
Replace the top portion of the inlet and tighten the knurled nut until you
cannot tighten it further. Restore the GC to normal operating conditions.
6850 Series II GC User Information
Routine Maintenance: Inlets
Maintaining a purged packed inlet
page 143 of
256