Removing a syringe
1.
2.
3.
Caution
Be careful not to bend the syringe needle. Only pull the syringe out of the car-
riage until clear. The needle bends easily when still seated in the needle
support.
4.
Replacing the needle support assembly
To remove the needle support assembly
1.
2.
3.
Released: March 2004
Disconnect the injector or sampler cable from the GC and lay the injector
or sampler on its back on a flat surface.
Loosen the plunger screw and raise the plunger carrier loop off the
syringe.
Open the syringe latch.
Carefully lift the syringe flange out of the flange guide until clear, then lift
the syringe needle out of the needle support.
Remove all vials from the turret, disconnect the injector or sampler cable
from the GC and lay the injector or liquid sampler on its back on a flat
surface.
Open the door.
Remove the syringe.
6850 Series II GC User Information
Automatic Liquid Sampler and Injector
(Figure
44):
Syringes