Site Prep Summary; Summary Of Gc Benchtop And Lab Requirements - Agilent Technologies 6850 Series II User Information

Gas chromatograph
Hide thumbs Also See for 6850 Series II:
Table of Contents

Advertisement

Site Prep Summary

Summary of GC benchtop and lab requirements

Table 39.
Physical Dimensions
Height
50 cm (19.7-inch)
61 cm (23-inch), FPD
Table 40.
Power Requirements and Heat Output
Line voltage (±10%)
Regular oven
100 V* (Japan)
120 V* (USA)
230 V* (Europe)
Fast-heating oven
120 V* (USA)
200 V* (Japan)
230 V* (Europe)
*Requires an isolated ground and dedicated outlet.
Table 41.
Temperature and Humidity Ranges
Temperature range
Operating
15–35°C
Maximum
5–40°C
Released: March 2004
Width
29 cm (11.4 inch), standard
34 cm (13.4 inch), cryo cooling
37 cm (14.6 inch), 6850 ALS
Current
Maximum power consumption
15 A
1440 W
15 A
1440 W
12 A maximum
2000 W
8 A minimum
20 A
2400 W
12 A
2400 W
11 A
2400 W
Humidity range
5–95%
5–95%
6850 Series II GC User Information
Summary of GC benchtop and lab requirements
Depth
54 cm (21.3-inch)
64 cm (25.2-inch), auxiliary
EPC
Line frequency
47.5 – 63 Hz
47.5 – 63 Hz
47.5 – 63 Hz
47.5 – 63 Hz
47.5 – 63 Hz
47.5 – 63 Hz
Altitude range
2000 m
Site Prep Summary
Weight
< 23 kg (51 lb.)
Heat output
4800 BTU/hour
4800 BTU/hour
4800 BTU/hour
4800 BTU/hour
4800 BTU/hour
4800 BTU/hour
page 254 of
256

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents