4.
Installing columns in the FID detector
Be sure you have the correct jet installed in your detector before installing a
column.
If the capillary column adapter is not installed on the detector, begin at step 1.
If it is installed, begin at step 5.
1.
2.
Released: March 2004
Wipe the column walls with a tissue dampened with isopropanol to
remove fingerprints and dust.
Assemble a brass nut and graphitized Vespel ferrule on the liner/adapter.
1/4-in. ferrule
Capillary column adapter
Insert the adapter straight into the detector base as far as possible.
Detector fitting
Capillary column adapter
6850 Series II GC User Information
1/4-in. nut
Columns and Traps
Installing capillary columns