FAULT 242
FAULT 243
FAULT 244
Possible cause:
FAULT 245
FAULT 247
FAULT 249
Possible cause:
FAULT 250
FAULT 252
FAULT 254
Possible causes:
FAULT 255
FAULT 257
Possible cause:
FAULT 259
Possible cause:
FAULT 262
FAULT 263
FAULT 264
Possible causes:
Released: March 2004
Aux3 faulty fact cal
Aux4 faulty fact cal
Aux5 faulty fact cal
Defective auxiliary flow module
The auxiliary flow module's calibration is invalid.
Detector module rev
Inlet module rev
Aux module revision
Flow module version is obsolete and is not compatible with the GC
The flow module's EEPROM is not recognized by the GC. The module is
unusable. The version of the module is not compatible with the version of the
GC.
Wrong det module
Wrong inlet module
Wrong aux module
Wrong flow module installed for detector/inlet/aux
Defective detector/inlet/aux flow module
Flow module revision is obsolete and not recognized
The flow module installed for the detector/inlet/aux is unusable. The module is
not recognized as a valid detector flow module.
Det invalid type
Inlet invalid type
The installed detector/inlet flow module is not supported by the GC
The detector/inlet flow module revision is not supported by the firmware
installed in the GC.
Det type mismatch
GC detector type was changed but its board was not replaced or the reverse
The detector flow module type does not match the installed detector
electronics board.
Front RS232 failed
Back RS232 failed
Sampler RS232 failed
Defective main board
The RS232 communications on the main board are faulty. Communication
between the GC and the other device is halted.
6850 Series II GC User Information
Messages
Fault messages
page 97 of
256