Align Mode light is on, all others are flashing
(6850 Automatic Liquid Sampler only)
All lights are on
Released: March 2004
Probable cause
Turret alignment is in progress.
Suggested action
Install the calibration probe and press the Align button to complete the
alignment.
Probable cause
The cables connections are loose
There is a board failure
There is a firmware revision conflict
Suggested action
1.
Check all cable connections.
2.
Turn the instrument off, then on again.
3.
If the lights remain on, obtain Agilent service.
6850 Series II GC User Information
Automatic Liquid Sampler and Injector
page 241 of